Testing Piezometers for Functionality Prior to Installation
Figure 1. Soil Instruments Standard VW Piezometer
1. Introduction
Upon
receiving your Piezometers you should visually inspect your shipment to
ensure that no damage has occured in transit.
When leaving Soil
Instruments our Piezometers are fully tested to ensure they are
functioning correctly however we do recommend testing the Piezometers
prior to installation.
This includes taking a base reading on site along
with testing during the installation process.
2. Testing the Functionality of the Piezometers
Before
taking the Piezometers to site a simple functionality test can be
performed to ensure the Piezometers are functioning correctly.
This can
be done simply by:
Filling a bucket or an equivalent with water
Connecting
the Piezometer to a logger or readout unit and ensure you are getting a
stable reading. If you are receiving an unstable reading you may need
to cycle through the sweep ranges on the logger to find the correct
sweep range. There is also a knowledgebase article on our website that
details which sweep range is required for different vibrating wire
instruments.
Once you are receiving a stable reading please make note of this reading.
To
then test the functionality please lower the Piezometer into the water
at measured intervals (see Figure 2.).
You
will then need to calculate the values you have into mH20 which are
then comparable to the intervals you have measured. The calculations for
this conversion can be found in the data reduction section of the W9
Piezometer manual.
Figure 2. Lowering a Piezometer into a body of water for testing
3. Possible differences seen
When
conducting these tests you are looking for a stable linear response
from the Piezometer showing increasing pressure as the Piezometer is
inserted deeper into the water.
The readings you receive from the piezometer should typically fall within
25 F^2/1000 digits of the values stated on the Piezometers calibration
sheet. We say typically 25 F^2/1000 digits as there is a
possibility that transital influences may affect the readings you are
receiving slightly, however this will not cause any reduction of the
Piezometers accuracy.
Please remember to compensate for any barometric
differences that you may be experiencing compared to when the Piezometer
was calibrated, calculations for this can be found in the instruments
manual.
4. Functionality tests on Site
We
strongly recommend further tests of the Piezometers on site including
taking your base reading.
If this base reading on site is not taken then
once installed you will only be able to measure pressure change from
when the Piezometer is first entered into the ground.
We also recommend
testing the Piezometer throughout the installation process to ensure
that no damage has occurred to the Piezometer.
If
your Piezometers readings fall outside the parameters stated above
please submit a support Email detailing the differences you are
experiencing.